Document Outline
- Introduction
- Scope and Application
- Summary of Method
- Definitions
- Contamination and Interferences
- Safety
- Equipment and Supplies
- Reagents and Standards
- Sample Collection, Preservation, and Storage
- Quality Control
- Calibration and Standardization
- Procedure
- Data Analysis and Calculations
- Method Performance
- Pollution Prevention
- Waste Management
- References
- Tables and Diagrams
- Glossary
- Appendix A
Method 1630
Methyl Mercury in Water by Distillation, Aqueous Ethylation, Purge and Trap, and Cold Vapor Atomic Fluorescence Spectrometry
August, 1998
U.S. Environmental Protection Agency
Office of Water
Office of Science and Technology
Engineering and Analysis Division (4303)
401 M Street SW
Washington, D.C. 20460
Acknowledgments
This method was prepared under the direction of William A. Telliard of the Engineering and Analysis
Division (EAD) within the U.S. Environmental Agency's (EPA's) Office of Science and Technology (OST).
The method was prepared by Nicholas Bloom of Frontier Geosciences under EPA Contract 68-C3-0337
with the DynCorp Environmental Programs Division. Additional assistance in preparing the method was
provided by DynCorp Environmental and Interface, Inc.
Disclaimer
This draft method has been reviewed and approved for publication by the Analytical Methods Staff within
the Engineering and Analysis Division of the U.S. Environmental Protection Agency. Mention of trade
names or commercial products does not constitute endorsement or recommendation for use. EPA plans
further validation of this draft method. The method may be revised following validation to reflect results of
the study.
EPA welcomes suggestions for improvement of this method. Suggestions and questions concerning this
method or its application should be addressed to:
W.A. Telliard
Engineering and Analysis Division (4303)
U.S. Environmental Protection Agency
401 M Street SW
Washington, D.C. 20460
Phone: 202/260-7134
Fax: 202/260-7185
Introduction
This analytical method supports water quality monitoring programs authorized under the Clean Water Act (CWA, the "Act"). CWA Section 304(a) requires EPA to publish water quality criteria that reflect the latest scientific knowledge
concerning the physical fate (e.g., concentration and dispersal) of pollutants, the effects of pollutants on ecological and
human health, and the effect of pollutants on biological community diversity, productivity, and stability.
CWA Section 303 requires each state to set a water quality standard for each body of water within its boundaries. A
state water quality standard consists of a designated use or uses of a water body or a segment of a water body, the water
quality criteria that are necessary to protect the designated use or uses, and an antidegradation policy. These water
quality standards serve two purposes: (1) they establish the water quality goals for a specific water body, and (2) they
are the basis for establishing water quality-based treatment controls and strategies beyond the technology-based controls
required by CWA Sections 301(b) and 306.
In defining water quality standards, the state may use narrative criteria, numeric criteria, or both. However, the 1987
amendments to CWA required states to adopt numeric criteria for toxic pollutants (designated in Section 307(a) of the
Act) based on EPA Section 304(a) criteria or other scientific data, when the discharge or presence of those toxic
pollutants could reasonably be expected to interfere with designated uses.
In some cases, these water quality criteria are as much as 280 times lower than those achievable using existing EPA
methods and required to support technology-based permits. Therefore, EPA developed new sampling and analysis
methods to specifically address state needs for measuring toxic metals at water quality criteria levels, when such
measurements are necessary to protect designated uses in state water quality standards. The latest criteria published
by EPA are those listed in the National Toxics Rule (58 FR 60848) and the Stay of Federal Water Quality Criteria for
Metals (60 FR 22228). These rules include water quality criteria for 13 metals, and it is these criteria on which the
new sampling and analysis methods are based. Method 1630 was specifically developed to provide reliable
measurements of methyl mercury at EPA WQC levels.
In developing methods for determination of trace metals, EPA found that one of the greatest difficulties was precluding
sample contamination during collection, transport, and analysis. The degree of difficulty, however, is highly dependent
on the metal and site-specific conditions. This method is designed to preclude contamination in nearly all situations.
It also contains procedures necessary to produce reliable results at the lowest ambient water quality criteria published
by EPA. In recognition of the variety of situations to which this method may be applied, and in recognition of
continuing technological advances, Method 1630 is performance based. Alternative procedures may be used so long
as those procedures are demonstrated to yield reliable results.
Requests for additional copies of this method should be directed to:
U.S. EPA NCEPI
11209 Kenwood Road
Cincinnati, OH 45242
513/489-8190
Method 1630
Note: This method is performance based. The laboratory is permitted to omit any step or modify any
procedure provided that all performance requirements in this method are met. The laboratory may not
omit any quality control analyses. The terms “shall,” “must,” and “may not” define procedures required
for producing reliable data at water quality criteria levels. The terms “should” and “may” indicate
optional steps that may be modified or omitted if the laboratory can demonstrate that the modified
method produces results equivalent or superior to results produced by this method.
1.0 Scope and Application
1.1 This method is for determination of methyl mercury (CH3Hg3) in filtered and unfiltered water by
distillation, aqueous ethylation, purge and trap, desorption, and cold vapor atomic fluorescence
spectrometry (CVAFS). This method is for use in EPA's data gathering and monitoring programs
associated with the Clean Water Act, the Resource Conservation and Recovery Act, the
Comprehensive Environmental Response, Compensation and Liability Act, and the Safe Drinking
Water Act. The method is based on a contractor-developed method (Reference 1) and on
peer-reviewed, published procedures for the determination of CH3Hg in aqueous samples, ranging
from sea water to sewage effluent (References 2-7).
1.2 This method is accompanied by Method 1669: Sampling Ambient Water for Determination of
Trace Metals at EPA Water Quality Criteria Levels (Sampling Method). The Sampling Method is
necessary to preclude contamination during the sampling process.
1.3 This method is designed for determination of CH3Hg3 in the range of 0.02-5 ng/L and may be
extended to higher levels by selection of a smaller sample size.
1.4 The ease of contaminating ambient water samples with the metal(s) of interest and interfering
substances cannot be overemphasized. This method includes suggestions for improvements in
facilities and analytical techniques that should maximize the ability of the laboratory to make
reliable trace metal determinations and minimize contamination (Section 4.0).
1.5 The detection limit and minimum level of quantitation in this method are usually dependent on the
level of background elements rather than instrumental limitations. The method detection limit
(MDL; 40 CFR 136, Appendix B) for CH3Hg has been determined to be 0.02 ng/L when no
background elements or interferences are present. The minimum level (ML) has been established
as 0.06 ng/L. An MDL as low as 0.009 ng/L can be achieved for low CH3Hg samples by using
extra caution in sample handling and reagent selection, particularly the use of “for ultra-low level
only” distillation equipment.
1.6 Clean and ultraclean—The terms "clean" and "ultraclean" have been applied to the techniques
needed to reduce or eliminate contamination in trace metal determinations. These terms are not
used in this method because they lack an exact definition. However, the information provided in
this method is consistent with the summary guidance on clean and ultraclean techniques.
1.7 This method follows the EPA Environmental Methods Management Council's "Format for Method
Documentation."
1.8 This method is "performance based." The analyst is permitted to modify the method to overcome
interferences or lower the cost of measurements if all performance criteria are met. Section 9.1.2
gives the requirements for establishing method equivalency.
1.9 Any modification of this method, beyond those expressly permitted, shall be considered a major
modification subject to application and approval of alternate test procedures under 40 CFR 136.4
and 136.5.
1.10 This method should be used only by analysts who are experienced in the use of CVAFS techniques
and who are thoroughly trained in the sample handling and instrumental techniques described in
this method. Each analyst who uses this method must demonstrate the ability to generate
acceptable results using the procedure in Section 9.2.
1.11 This method is accompanied by a data verification and validation guidance document, Guidance on
the Documentation and Evaluation of Trace Metals Data Collected for CWA Compliance
Monitoring. Data users should state data quality objectives (DQOs) required for a project before
this method is used.
2.0 Summary of Method
2.1 A 100-2000 mL sample is collected directly into specially cleaned, pretested, fluoropolymer or
borosilicate bottle(s) using sample handling techniques specially designed for collection of metals
at trace levels (Reference 6).
2.2 For dissolved CH3Hg, samples are filtered through a 0.45-µm capsule filter.
2.3 Fresh water samples are preserved by adding 4 mL/L of pretested 11.6 M HCl, while saline
samples ([Cl-] > 500 ppm) are preserved with 2 mL/L of 9 M H2SO4solution, to avoid distillation
interferences caused by excess chloride.
2.4 Prior to analysis, a 45-mL sample aliquot is placed in a specially designed fluoropolymer
distillation vessel, and 35 mL of the water is distilled into the receiving vessel at 125°C under N2 flow.
2.5 After distillation, the sample is adjusted to pH 4.9 with an acetate buffer and ethylated in a closed
purge vessel by the addition of sodium tetraethyl borate (NaBEt ).
2.6 The ethyl analog of CH3Hg, methylethyl mercury (CH3CH3CH2Hg), is separated from solution by
purging with N2 onto a graphitic carbon (Carbotrap®) trap.
2.7 The trapped methylethyl mercury is thermally desorbed from the Carbotrap® trap into an inert gas
stream that carries the released methylethyl mercury first through a pyrolytic decomposition
column, which converts organo mercury forms to elemental mercury (Hg0), and then into the cell of
a cold-vapor atomic fluorescence spectrometer (CVAFS) for detection.
2.8 Quality is ensured through calibration and testing of the distillation, ethylation, purging, and
detection systems.
3.0 Definitions
3.1 Apparatus: Throughout this method, the sample containers, sampling devices, instrumentation,
and all other materials and devices used in sample collection, sample processing, and sample
analysis that come in contact with the sample and therefore require careful cleaning will be referred
to collectively as the Apparatus.
3.2 Dissolved methyl mercury: All distillable CH3Hg forms and species found in the filtrate of an
aqueous solution that has been filtered through a 0.45 micron filter.
3.3 Methyl mercury: All acid-distillable Hg, which, upon reaction with NaBEt4 yields methylethyl
mercury. This includes, but is not limited to, CH3Hg+, strongly organo-complexed CH3Hg
compounds, adsorbed particulate CH3Hg, and CH3Hg bound in microorganisms. In freshly
collected samples, dimethyl mercury ((CH3) Hg) will not be recovered as CH3Hg, but in samples
which have been acidified for several days, most (CH3)2Hg has broken down to CH3Hg. In this
method, CH3Hg and total recoverable CH3Hg are synonymous.
3.4 Definitions of other terms used in this method are given in the glossary at the end of the method.
4.0 Contamination and Interferences
4.1 Preventing ambient water samples from becoming contaminated during the sampling and analysis
process constitutes one of the greatest difficulties encountered in trace metals determinations. Over
the last two decades, marine chemists have come to recognize that much of the historical data on
the concentrations of dissolved trace metals in seawater are erroneously high because the
concentrations reflect contamination from sampling and analysis rather than ambient levels.
Therefore, it is imperative that extreme care be taken to avoid contamination when collecting and
analyzing ambient water samples for trace metals.
4.2 Samples may become contaminated by numerous routes. Potential sources of trace metal
contamination during sampling include: metallic or metal-containing labware (e.g., talc gloves that
contain high levels of zinc), containers, sampling equipment, reagents, and reagent water;
improperly cleaned and stored equipment, labware, and reagents; and atmospheric inputs such as
dirt and dust. Even human contact can be a source of trace metal contamination. For example, it
has been demonstrated that dental work (e.g., mercury amalgam fillings) in the mouths of
laboratory personnel can contaminate samples that are directly exposed to exhalation (Reference
5).
4.3 Contamination control
4.3.1 Philosophy—The philosophy behind contamination control is to ensure that any object or
substance that contacts the sample is metal free and free from any material that may
contain Hg or CH3Hg.
4.3.1.1 The integrity of the results produced cannot be compromised by
contamination of samples. This method and the Sampling Method give
requirements and suggestions for control of sample contamination.
4.3.1.2 Substances in a sample cannot be allowed to contaminate the laboratory
work area or instrumentation used for trace metals measurements. This
method gives requirements and suggestions for protecting the laboratory.
4.3.1.3 Although contamination control is essential, personnel health and safety
remain the highest priority. The Sampling Method and Section 5 of this
method give requirements and suggestions for personnel safety.
4.3.2 Avoid contamination—The best way to control contamination is to completely avoid
exposure of the sample to contamination in the first place. Avoiding exposure means
performing operations in an area known to be free from contamination. Two of the most
important factors in avoiding/reducing sample contamination are (1) an awareness of
potential sources of contamination and (2) strict attention to the work being done.
Therefore, it is imperative that the procedures described in this method be carried out by
well-trained, experienced personnel.
4.3.3 Use a clean environment—The ideal environment for processing samples is a class 100
clean room. If a clean room is not available, all sample preparation should be performed
in a class 100 clean bench or a nonmetal glove box fed by mercury-free and particle-free
air or nitrogen. Digestions should be performed in a nonmetal fume hood situated, ideally
in the clean room.
4.3.4 Minimize exposure—The Apparatus that will contact samples, blanks, or standard
solutions should be opened or exposed only in a clean room, clean bench, or glove box so
that exposure to an uncontrolled atmosphere is minimized. When not in use, the
Apparatus should be covered with clean plastic wrap, stored in the clean bench or in a
plastic box or glove box, or bagged in clean zip-type bags. Minimizing the time between
cleaning and use will also minimize contamination.
4.3.5 Clean work surfaces—Before a given batch of samples is processed, all work surfaces in
the hood, clean bench, or glove box in which the samples will be processed should be
cleaned by wiping with a lint-free cloth or wipe soaked with reagent water.
4.3.6 Wear gloves—Sampling personnel must wear clean, non talc gloves during all operations
involving handling of the Apparatus, samples, and blanks. Only clean gloves may touch
the Apparatus. If another object or substance is touched, the glove(s) must be changed
before again handling the Apparatus. If it is even suspected that gloves have become
contaminated, work must be halted, the contaminated gloves removed, and a new pair of
clean gloves put on. Wearing multiple layers of clean gloves will allow the old pair to be
quickly stripped with minimal disruption to the work activity.
4.3.7 Use metal-free Apparatus—All Apparatus used for determination of CH3Hg at ambient
water quality criteria levels must be nonmetallic and free of material that may contain
metals.
4.3.7.1 Construction materials—Only fluoropolymer or borosilicate glass
containers should be used for samples that will be analyzed for Hg
because Hg vapors can diffuse in or out of other materials, resulting in
results that are biased low or high. All materials, regardless of
construction, that will directly or indirectly contact the sample must be
cleaned using the procedures in this method and must be known to be
clean and mercury free before proceeding.
4.3.7.2 Serialization—It is recommended that serial numbers be indelibly marked
or etched on each piece of Apparatus so that contamination can be traced,
and logbooks should be maintained to track the sample from the container
through the labware to introduction into the instrument. It may be useful
to dedicate separate sets of labware to different sample types; e.g.,
receiving waters vs. effluents. However, the Apparatus used for
processing blanks and standards must be mixed with the Apparatus used
to process samples so that contamination of all labware can be detected.
4.3.7.3 The laboratory or cleaning facility is responsible for cleaning the
Apparatus used by the sampling team. If there are any indications that the Apparatus is not clean when received by the sampling team (e.g., ripped
storage bags), an assessment of the likelihood of contamination must be
made. Sampling must not proceed if it is possible that the Apparatus is contaminated. If the Apparatus is contaminated, it must be returned to the
laboratory or cleaning facility for proper cleaning before any sampling
activity resumes.
4.3.8 Avoid sources of contamination—Avoid contamination by being aware of potential
sources and routes of contamination.
4.3.8.1 Contamination by carryover—Contamination may occur when a sample
containing a low concentration of CH3Hg is processed immediately after a
sample containing a relatively high concentration. When an unusually
concentrated sample is encountered, a ethylation blank should be analyzed
immediately following the sample to check for carryover. Samples known
or suspected to contain the lowest concentration of CH3Hg should be
analyzed first followed by samples containing higher levels.
4.3.8.2 Contamination by samples—Significant laboratory or instrument
contamination may result when untreated effluents, in-process waters,
landfill leachates, and other samples containing high concentrations of Hg
or CH3Hg are processed and analyzed. This method is not intended for
application to these samples, and samples containing high concentrations
of trace metals should not be permitted into the clean room and laboratory
dedicated for processing trace metals samples.
4.3.8.3 Contamination by indirect contact—Apparatus that may not directly come
in contact with the samples may still be a source of contamination. For
example, clean tubing placed in a dirty plastic bag may pick up
contamination from the bag and subsequently transfer the contamination
to the sample. Therefore, it is imperative that every piece of the
Apparatus that is directly or indirectly used in the collection, processing,
and analysis of samples be thoroughly cleaned (see Section 6.1.2).
4.3.8.4 Contamination by airborne particulate matter—Airborne particles are less
obvious substances capable of contaminating samples. Samples may be
contaminated by airborne dust, dirt, particles, or vapors from unfiltered
air supplies; nearby corroded or rusted pipes, wires, or other fixtures; or
metal-containing paint. Whenever possible, sample processing and
analysis should occur as far as possible from sources of airborne
contamination.
4.4 Interferences
4.4.1 When the method is properly applied, no significant interferences have been observed in
the analysis of ambient waters.
4.4.2 Distillation of CH3Hg from solution requires a carefully controlled level of HCl in
solution. Distillation will not be quantitative if too little HCl is added, but too much HCl
results in co-distillation of HCl fumes, which interfere with the ethylation procedure.
Therefore fresh water samples must be preserved only with between 0.3% and 0.5% (v/v)
11.6 M HCl, and salt water samples with between 0.1% and 0.2% (v/v) 9 M H2SO4.
4.4.3 Samples preserved with nitric acid (HNO3) cannot be analyzed for CH3Hg as the analyte is
partially decomposed in the distillation step by this reagent.
4.4.4 The fluorescent intensity is strongly dependent upon the presence of molecular species in
the carrier gas that can cause "quenching" of the excited atoms. The Carbotrap® trap
eliminates quenching due to trace gases, but it still remains the analyst's responsibility to
ensure high purity inert carrier gas and a leak-free analytical train. In some rare cases
(such as oil polluted water) low molecular weight organic compounds may purge with the
methylethyl mercury and collect on the Carbotrap® trap, subsequently resulting in signal
quenching during elution. Such cases are best treated by sample dilution prior to
distillation.
4.4.5 Recent investigations have shown that a positive artifact is possible with the distillation
procedure in cases where high inorganic Hg concentrations are present (Reference 7). In
natural waters, approximately 0.01 to 0.05% of the ambient inorganic Hg in solution may
be methylated by ambient organic matter during the distillation step. In most waters,
where the percent CH3Hg is 1-30% of the total, this effect is trivial. However, the analyst
should be aware that in inorganic Hg contaminated waters, the fraction CH3Hg can be <
1% of the total, and so flagging of the data (as representing a maximum estimate of
CH3Hg concentration) may be warranted. In samples with high levels of divalent mercury
(Hg(II)), solvent extraction may be preferable to distillation (Reference 7).
5.0 Safety
5.1 The toxicity or carcinogenicity of each chemical used in this method has not been precisely
determined; however, each compound should be treated as a potential health hazard. Exposure to
these compounds should be reduced to the lowest possible level. It is suggested that the laboratory
perform personal hygiene monitoring of each analyst using this method and that the results of this
monitoring be made available to the analyst.
5.1.1 Chronic Hg exposure may cause kidney damage, muscle tremors, spasms, personality
changes, depression, irritability, and nervousness. Organo-mercurials may cause
permanent brain damage. Because of the toxicological and physical properties of CH3Hg,
pure standards should be handled only by highly trained personnel thoroughly familiar with
handling and cautionary procedures and the associated risks.
5.1.2 It is recommended that the laboratory purchase a dilute standard solution of CH3Hg for
this method. If primary solutions are prepared, they shall be prepared in a hood, and a
NIOSH/MESA-approved toxic gas respirator shall be worn when high concentrations are
handled.
5.2 This method does not address all safety issues associated with its use. The laboratory is
responsible for maintaining a current awareness file of OSHA regulations for the safe handling of
the chemicals specified in this method. A reference file of material safety data sheets (MSDSs)
should also be made available to all personnel involved in these analyses. Additional information
on laboratory safety can be found in References 7-10. The references and bibliography at the end
of Reference 10 are particularly comprehensive in dealing with the general subject of laboratory
safety.
5.3 Samples suspected of containing high concentrations of CH3Hg are handled using essentially the
same techniques employed in handling radioactive or infectious materials. Well-ventilated,
controlled access laboratories are required. Assistance in evaluating the health hazards of
particular laboratory conditions may be obtained from certain consulting laboratories and from
State Departments of Health or Labor, many of which have an industrial health service. Each
laboratory must develop a strict safety program for handling CH3Hg.
5.3.1 Facility—When samples known or suspected to contain high concentrations of CH Hg are
handled, all operations (including removal of samples from sample containers, weighing,
transferring, and mixing) should be performed in a glove box demonstrated to be leakproof
or in a fume hood demonstrated to have adequate airflow. Gross losses to the laboratory
ventilation system must not be allowed. Handling of the dilute solutions normally used in
analytical and animal work presents no inhalation hazard except in an accident.
5.3.2 Protective equipment—Disposable plastic gloves, apron or lab coat, safety glasses or
mask, and a glove box or fume hood adequate for radioactive work should be used.
During analytical operations that may give rise to aerosols or dusts, personnel should wear
respirators equipped with activated carbon filters.
5.3.3 Training—Workers must be trained in the proper method of removing contaminated gloves
and clothing without contacting the exterior surfaces.
5.3.4 Personal hygiene—Hands and forearms should be washed thoroughly after each
manipulation and before breaks (coffee, lunch, and shift).
5.3.5 Confinement—Isolated work areas posted with signs, segregated glassware and tools, and
plastic absorbent paper on bench tops will aid in confining contamination.
5.3.6 Effluent vapors—The effluent from the CVAFS should pass through either a column of
activated charcoal or a trap containing gold or sulfur to amalgamate or react with Hg
vapors.
5.3.7 Waste handling—Good technique includes minimizing contaminated waste. Plastic bag
liners should be used in waste cans. Janitors and other personnel must be trained in the
safe handling of waste.
5.3.8 Decontamination
5.3.8.1 Decontamination of personnel—Use any mild soap with plenty of
scrubbing action.
5.3.8.2 Glassware, tools, and surfaces—Activated carbon powder will adsorb
CH3Hg, eliminating the possible volatilization of CH3Hg. Satisfactory
cleaning may be accomplished by dusting a surface lightly with activated
carbon powder, then washing with any detergent and water.
5.3.9 Laundry—Clothing known to be contaminated should be collected in plastic bags. Persons
who convey the bags and launder the clothing should be advised of the hazard and trained
in proper handling. If the launderer knows of the potential problem, the clothing may be
put into a washer without contact. The washer should be run through a cycle before being
used again for other clothing.
5.3.10 Wipe tests—A useful method of determining cleanliness of work surfaces and
tools is to wipe the surface with a piece of filter paper. Extraction and analysis by
this method can achieve a limit of detection of less than 1 ng per wipe. Less than
0.1 µg per wipe indicates acceptable cleanliness; anything higher warrants further
cleaning. More than 10 µg on a wipe constitutes an acute hazard, requires prompt
cleaning before further use of the equipment or work space, and indicates that
unacceptable work practices have been employed.
6.0 Equipment and Supplies
NOTE: The mention of trade names or commercial products in this method is for illustrative
purposes only and does not constitute endorsement or recommendation for use by the
Environmental Protection Agency. Equivalent performance may be achievable using apparatus,
materials, or cleaning procedures other than those suggested here. The laboratory is
responsible for demonstrating equivalent performance.
6.1 Sampling equipment
6.1.1 Sample collection bottles-fluoropolymer or borosilicate glass, 125- to 1000-mL, with
fluoropolymer or fluoropolymer-lined cap.
6.1.2 Cleaning
6.1.2.1 New bottles are cleaned by heating to 65-75°C in 4 N HCl for at least 48
h. The bottles are cooled, rinsed three times with reagent water, and filled
with reagent water containing 1% HCl. These bottles are capped and
placed in a clean oven at 60-70°C overnight. After cooling, they are
rinsed three more times with reagent water, filled with reagent water
containing 0.4% (v/v) HCl, capped, and placed in a mercury-free class
100 clean bench until the outside of the bottle is dry. The caps are then
tightened with a wrench and the bottles are double-bagged in new
polyethylene zip-type bags. The capped bottles are stored in wooden or
plastic boxes until use.
6.1.2.2 To avoid long-term accumulation of Hg or CH3Hg on the bottle walls due
to trace organic coatings, used bottles are filled with reagent water
containing 0.02 N BrCl solution and allowed to stand over night. The
BrCl is neutralized with the addition of 0.2 mL of 20% NH2OH solution.
The bottles are then cleaned exactly as in Section 6.1.2.1, except that they
soak only 6-12 h in hot 4 N HCl.
6.1.2.3 Bottle blanks should be analyzed as described in Section 9.4.4.1 to verify
the effectiveness of the cleaning procedures.
6.1.3 Filtration Apparatus
6.1.3.1 Filter—0.45-µm, 15-mm diameter capsule filter (Gelman Supor 12175, or
equivalent)
6.1.3.2 Peristaltic pump—115-V a.c., 12-V d.c., internal battery, variable-speed,
single-head (Cole-Parmer, portable, "Masterflex L/S," Catalog No. H-
07570-10 drive with Quick Load pump head, Catalog No. H-07021-24,
or equivalent).
6.1.3.3 Tubing—styrene/ethylene/butylene/silicone (SEBS) resin for use with
peristaltic pump, approx 3/8-in i.d. by approximately 3 ft (Cole-Parmer
size 18, Catalog No. G-06464-18, or approximately 1/4-in i.d., Cole-
Parmer size 17, Catalog No. G-06464-17, or equivalent). Tubing is
cleaned by soaking in 5-10% HCl solution for 8-24 h. It is rinsed with
reagent water on a clean bench in a clean room and dried on the clean
bench by purging with metal-free air or nitrogen. After drying, the tubing
is double-bagged in clear polyethylene bags, serialized with a unique
number, and stored until use.
6.2 Equipment for bottle and glassware cleaning
6.2.1 Vat, 100-200 L, high-density polyethylene (HDPE), half filled with 4 N HCl in reagent
water.
6.2.2 Panel immersion heater, 500-W, all-fluoropolymer coated, 120 vac (Cole-Parmer H-
03053-04, or equivalent)
NOTE: Read instructions carefully!! The heater will maintain steady state, without
temperature feedback control, of 60-75°C in a vat of the size described. However, the
equilibrium temperature will be higher (up to boiling) in a smaller vat. Also, the heater plate
MUST be maintained in a vertical position, completely submerged and away from the vat walls
to avoid melting the vat or burning out!
6.2.3 Laboratory sink in class 100 clean area, with high-flow reagent water (Section 7.1) for
rinsing.
6.2.4 Clean bench, class 100, for drying rinsed bottles.
6.2.5 Oven, stainless steel, in class 100 clean area, capable of maintaining ± 5°C in the 60-70°C
temperature range.
6.3 Cold vapor atomic fluorescence spectrometer (CVAFS): The CVAFS system used may either be
purchased from a supplier, or built in the laboratory from commercially available components.
6.3.1 Commercially available: Tekran Model 2357 CVAFS, Brooks-Rand Model III CVAFS,
or equivalent
6.3.2 Custom-built CVAFS (Reference 11). Figure 1 shows the schematic diagram. The system
consists of the following:
6.3.2.1 Low-pressure 4-W mercury vapor lamp
6.3.2.2 Far UV quartz flow-through fluorescence cell—12 mm x 12 mm x 45
mm, with a 10-mm path length (NSG Cells, or equivalent).
6.3.2.3 UV-visible photomultiplier (PMT)—sensitive to < 230 nm. This PMT is
isolated from outside light with a 253.7-nm interference filter (Oriel
Corp., or equivalent).
6.3.2.4 Photometer and PMT power supply (Oriel Corp., or equivalent), to
convert PMT output (nanoamp) to millivolts
6.3.2.5 Black anodized aluminum optical block—holds fluorescence cell, PMT,
and light source at perpendicular angles, and provides collimation of
incident and fluorescent beams (Frontier Geosciences Inc., or equivalent).
6.3.2.6 Flowmeter, with needle valve capable of keeping the carrier gas at a
reproducible flow rate of 30 mL/min
6.3.2.7 Ultra high-purity argon (grade 5.0)
6.4 Equipment for CH3Hg purging system—Figure 2a shows the schematic diagram for the purging
system. The system consists of the following:
6.4.1 Flow meter/needle valve—capable of controlling and measuring gas flow rate to the purge
vessel at 350 (± 50) mL/min.
6.4.2 Fluoropolymer fittings—connections between components and columns are made using
6.4-mm o.d. fluoropolymer tubing and fluoropolymer friction-fit or threaded tubing
connectors. Connections between components requiring mobility are made with 3.2-mm
o.d. fluoropolymer tubing because of its greater flexibility.
6.4.3 Cold vapor generator (bubbler)—200-mL borosilicate glass (15 cm high x 5.0 cm
diameter) with standard taper 24/40 neck, fitted with a sparging stopper having a coarse glass frit that extends to within 0.2 cm of the bubbler bottom (Frontier Geosciences, Inc.,
or equivalent).
6.5 Equipment for the isothermal gas chromatography (GC) system.
6.5.1 Figure 1 shows the schematic for the interface of the GC with the CVAFS detector
(Reference 6).
6.5.2 Figure 2b shows the orientation consideration for purging and desorbing CH3Hg from the
Carbotrap® traps.
6.5.3 Carbotrap® traps—10-cm x 6.5-mm o.d. x 4-mm i.d. quartz tubing. The tube is filled with
3.4 cm of 30/45 mesh Carbotrap® graphitic carbon adsorbant (Supelco, Inc., or
equivalent). The ends are plugged with silanized glass wool.
6.5.3.1 Traps are fitted with 6.5-mm i.d. fluoropolymer friction-fit sleeves for
making connection to the system. When traps are not in use,
fluoropolymer end plugs are inserted in trap ends to eliminate
contamination.
6.5.3.2 At least six traps are needed for efficient operation.
6.5.3.3 Because the direction of flow is important in this analysis, the crimped end
of the Carbotrap® trap will be referred to as “side A,” while the
uncrimped end will be referred to as “side B.”
6.5.4 Heating of Carbotrap® traps—To desorb CH3Hg collected on a trap, heat for 45 sec to
450-500°C (a barely visible red glow when the room is darkened) with a coil consisting of
75 cm of 24-gauge Nichrome wire at a potential of 16-20 vac. Potential is applied and
finely adjusted with an autotransformer.
6.5.5 Timer—The heating interval is controlled by a timer-activated 120-V outlet, into which the
heating coil autotransformer is plugged.
6.5.6 Isothermal GC—Consists of two parts, a custom fabricated packed GC column, and a
custom fabricated constant temperature oven.
6.5.6.1 The column is 1 m long, made from 0.25 inch OD by 4 mm ID
borosilicate glass GC column tubing. The column is formed into an 8 cm
diameter coil, with 15 cm straight extensions from each end. The column
is silanized, packed in the coiled portion with 60/80 mesh 15% OV-3 on
acid-washed Chromasorb W, and then conditioned under inert gas flow at
200°C. A column meeting these specifications may be custom fabricated
(Supelco Inc., or equivalent).
6.5.6.2 The GC oven consists of a 500-watt aluminum jacketed heating mantle,
fitted with a custom machined fluoropolymer lid (14 cm OD by 1 cm
thick). The lid is attached with stainless steel screws and contains three
threaded holes (0.25 inch female NPT) in a triangular pattern in the top.
The spacing of the holes conforms exactly to the spacing between the two
15 cm glass extensions of the GC column.
6.5.6.3 Fluoropolymer fittings, with 0.25-inch male NPT threads on the bottom
and 0.25-inch compression fittings on top, are placed into the threaded
holes. The GC column is secured into the oven by passing the glass
extensions through two of the fluoropolymer fittings, so that 3 cm of the
glass extensions protrude from the top, and tightening the compression
fittings. The fluoropolymer lid holding the GC column is then screwed to
the top of the oven.
6.5.6.4 Temperature feedback control (110 ± 2°C) is achieved through a
thermocouple temperature controller. The oven is plugged into the
controller and the thermocouple probe is inserted through the third
fluoropolymer fitting in the lid, such that the sensor is located near the
center of the GC coil.
6.5.6.5 Several research groups have successfully interfaced the
Carbotrap®/CVAFS system directly to a commercial gas chromatograph.
The use of capillary column GC will result in better peak separation,
although at higher cost.
6.5.7 Pyrolytic column—The output from the GC oven is connected directly to a high
temperature column to decompose eluted organo-mercurial compounds to Hg0. The output
of the pyrolytic column is connected to the inlet of the CVAFS system.
6.5.7.1 The column consists of a 20-cm length of quartz tubing, packed over the
central 10 cm with quartz wool.
6.5.7.2 The column is heated to orange heat (~ 700°C) by a 1 m length of 22
gauge Nichrome wire, tightly wrapped around the quartz wool packed
portion of the tube. The temperature of the coil is adjusted by visual
inspection of the color, using a 0-120 volt autotransformer.
6.6 Recorder—Any multi-range millivolt chart recorder or integrator with a range compatible with the
CVAFS is acceptable. By using a two pen recorder with pen sensitivity offset by a factor of 10,
the dynamic range of the system is extended to 103.
6.7 Distillation unit—The distillation unit is a custom made temperature controlled aluminum block
heater, as shown schematically in Figure 3 (Frontier Geosciences Inc., or equivalent).
6.7.1 Heating block insulation—Each heating block is encased first in refractory spun rock
insulation (1 inch thickness) and then an exterior wood shell for rigidity.
6.7.2 Each heating block (10 cm wide x 20 cm long x 15 cm high) is bored with five 31 mm
diameter holes (evenly spaced), 120 mm deep. A 3/8 inch diameter hole is bored to 90%
of the block length, perpendicular to and behind the distillation tube holes, to accommodate
a cylindrical heating element. A 2 mm diameter hole is bored parallel to the heating
element hole, and 2 cm above it, to accommodate the temperature sensor.
6.7.3 Heating element—Each heating block is equipped with a 750 watt cylindrical heating
element, 6 inches long by 3/8 inch diameter (Omega Inc.), immobilized in its respective
hole by a dab of silicone glue.
6.7.4 Type J thermocouple probe—Each heating block is equipped with a type J thermocouple
probe immobilized in its respective hole by a dab of silicone glue.
6.7.5 Digital temperature controller—The heating element and thermocouple are connected to a
digital temperature controller.
6.7.6 Fluoropolymer vials with caps—The distillation unit is designed to accommodate 60-mL
fluoropolymer vials (part number 0202, Savillex, or equivalent). The original caps are
used to close the vials when distillate is to be stored until analysis.
6.7.6.1 For each distillation, two identical vials are needed: a distillation vessel
and a receiving vessel. For convenience, each vial should be engraved
with a line at 40.0 mL (obtained by weighing 40 g of water in the vial),
and a unique identification number, both on the vial and the cap.
6.7.6.2 Fluoropolymer vials are acid cleaned initially as described for other
fluoropolymer ware and stored filled with 0.5% HCl. After use, receiving
vials are rinsed with reagent water and filled with 0.5% HCL. The tubing
is looped around the cap as described in Section 6.7.7.1, and the vials are
placed in a 70°C (± 5°C) oven overnight. Cleaning is the same for the
distillation vials, with the exception that first the vials, caps, and tubing are thoroughly scrubbed with an alkaline detergent and test tube brush to
remove any residues from the samples.
6.7.7 Purge caps—The standard caps on the fluoropolymer vials are replaced with purge caps
(part number 33-2-2, Savillex, or equivalent) for distillation purposes.
6.7.7.1 Fluoropolymer tubing—each purge cap is threaded with a piece of 1/8
inch fluoropolymer tubing, approximately 30-40 cm long. One end is
pulled through one of the holes in the cap, down to a length that will allow
it to reach the bottom of the distillation vial when the vial is screwed onto
the cap. The bottom end of this tubing is cut at a 45° angle. The outside
end of the tubing is cut perpendicularly and is looped around and inserted
into the second cap hole when not in use (to keep the system closed and
clean).
6.7.8 Aluminum distillation cover—The cover for the heating block consists of a 5 cm high
|